index
productc details
work process
technical specification
contact us
Visual inspection of the cervix with acetic acid for cervical intraepithelial lesions A. Goel*,
G. Gandhi, S. Batra,
S. Bhambhani, V. Zutshi, P. Sachdeva
...MORE


 
Output Characteristics
Ligasealer Modes (At 475 KHz) Maximum Open Circuit Voltage Vpp (Vp) Maximum Short Circuit Current Arms Maximum Power Setting Watts Crest Factor
Low 340 (170) 2.1 95 1.5
Medimum 770 (400) 2.1 125 1.5
High 580 (295) 4.2 150 1.5
  Weight and Dimensions
Input Power Requirments
  Height : 4.5 inch (11.5 cm) Volats : 170-270 VAC
  Width : 14.4 inch (36 cm) Amps : 6A
  Length : 17.8 inch (45 cm) Input : 1.5 KW, 50/60 Hz
  Weight : 12 Kg  
   
Low Frequency Leakage (50-60Hz) Audio Volume
  Patient Leakage < 150 µArms   The mode indicator tones are adjustable to a minimum level of 45 dba at 1 meter.
Cooling Single speed fan
High Frequency Leakage Operating mode
Less than 150 mA rms., all output int 10s/30s
CERTIFIED TO IEC -60601-2-2 AND IS-7583
OUTPUT CHARACTERISTICS
Electro Surgical Unit Mode Output Freq. (kHz) AMximum P-P Volatage Rated Load (ohms) Max. Power (Watts) Crest Factor
Bipolar Micro/Coag Macro/Cut VLS 400
400
400
480
480
T.D
100
100
T.D
90
95
T.D
2.5
1.5
T.D
Monopolar Cut Pure/Cut 1 Blent/Cut 2 Endo/Gastro Cut 500
500
500
1800
2600
1500
300
500
200
350
200
250
1.5
2.4
1.5
Monopolar Coag Soft/Desiccate Spray /Fulgurate Spray 450
450

450
2300
6000

9500
150
500

500
150
150

150
4.4
7.2

8.0
Crest Factor is an indicator of a waveforms' ability to coagulate bleeders without cutting effect.
For Clinical and Technical Details :
divlabsindia.com www.ligasealer.com www.quadrf.com
www.rfscalpal.com www.plasmaeffect.com  
 
 